Contact Us
Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:
425 Fortune Blvd #206, Milford MA 01757
(Hereinafter called the Organization) and hereby declares that Organization is accredited in accordance with the recognized International Standard:
This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (as outlined by the joint ISO-ILAC-IAF Communiqué dated April 2017):
Accreditation claims for such testing and/or calibration services shall only be made from addresses referenced within this certificate. This Accreditation is granted subject to the system rules governing the Accreditation referred to above, and the Organization hereby covenants with the Accreditation body’s duty to observe and comply with the said rules.
For PJLA:
Tracy Szerszen
President
Perry Johnson Laboratory Accreditation, Inc. (PJLA)
755 W. Big Beaver, Suite 1325
Troy, Michigan 48084
Initial Accreditation Date:
November 15, 2016
Issue Date:
May 19, 2023
Expiration Date:
July 31, 2025
Revision Date
October 28, 2023
Accreditation No.:
89989
Certificate No.:
L23-396-R1
The validity of this certificate is maintained through ongoing assessments based on a continuous accreditation cycle. The validity of this certificate should be confirmed through the PJLA website: www.pjlabs.com
425 Fortune Blvd #206 Milford, MA 01757
Contact Name: Hugh Stinson Phone: 508-872-4811
Accreditation is granted to the facility to perform the following calibrations:
Thermodynamic
MEASURED INSTRUMENT, QUANTITY OR GAUGE | RANGE OR NOMINAL DEVICE SIZE AS APPROPRIATE | CALIBRATION AND MEASUREMENT CAPABILITY EXPRESSED AS AN UNCERTAINTY (±) | CALIBRATION EQUIPMENT AND REFERENCE STANDARDS USED | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Hart 9009 Fluke 9150 Fluke 1586A Type S reference Type B reference RTD Reference High Temperature Furnance ASTM E220 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| High Temperature Furnance Fluke 1586A Type C reference ASTM E220 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Hart 9009 Fluke 9150 RTD Reference Fluke 1586A High Temperature Furnance ASTM E220 |
Issue 05/2023
This supplement is in conjunction with certificate #L23-396-R1
© 2024 Nanmac Corporation